Description
STAMMOPUR GR
Instrument basic cleaner
Concentrate for manual basic cleaning of medical and dental instruments, accessories and components of medical devices for use in an ultrasonic bath.
The preparation is used in the process of repairing instruments in the AEMP (reprocessing unit for medical devices) in clinics and medical and dental practices.
The application takes place outside the regular instrument reprocessing.
At a glance
- Concentrate
- Removes tarnish, metal oxides, rust, mineral residues such as lime, impurities burnt in by sterilisation, stains
- Use only in the insert vessel
- Biodegradable
- Not for regular instrument reprocessing
Material compatibility and impurities
STAMMOPUR GR has a high material compatibility.
Due to the formulation in the strongly acidic range, however, not suitable for acid-sensitive materials such as light metals and damaged chrome plating.
Removes
- Tarnish,
- Metal oxides (e.g. rust),
- mineral residues (e.g. lime,),
- Impurities burnt in by sterilisation
and stains
From the following materials
- Stainless steel,
- Precious metal,
- Titanium,
- Glass,
- Porcelain,
- Ceramics
and plastic.
Application
The exclusively professional users are trained specialists.
STAMMOPUR GR has a very high cleaning effect, which, however, only becomes effective in connection with the ultrasonic action in the ultrasonic bath and increased temperature (50-60 °C).
The application is carried out exclusively in an ultrasonic bath in an indirect process in insert vessels, e.g. plastic hanging basins or beakers.
For example, STAMMOPUR R, 2 % (medical) or TICKOMED 1, 1 % (dental) can be used as contact liquid in the ultrasonic bath.
Dosage and contact time
Dosage:
in ultrasonic bath: 5%
without ultrasound: -
Exposure time:
in ultrasonic bath: 3-10 min
without ultrasound: -
Temperature: 50-60 °C
Composition per 100g:
> 10 % Phosphoric acid,
1-< 10 % nichtionische Tenside,
1-< 10 % kationische Tenside
Produkt-pH-Bereich: stark sauer
Die Anwendung erfolgt verdünnt mit Wasser mit mindestens Trinkwasserqualität.
Eine Beimischung von Zusätzen in die Reinigungslösung ist nicht zulässig.
Besondere Hinweise
Optiken, Lichtleiter und geklebte Teile nicht mit Ultraschall behandeln.
Bohrer und weitere rotierende Kleininstrumente nicht mit STAMMOPUR GR behandeln.
Nicht medizinischer Anwendungsbereich:
STAMMOPUR GR kann zur Reinigung von Arbeitsschutzgeräten und Teilen von Tauchausrüstungen (Lungenautomaten) unter den gleichen Einsatzbedingungen wie beschrieben im Ultraschallbad eingesetzt werden.
Gefahrenhinweise
H314 Verursacht schwere Verätzungen der Haut und schwere Augenschäden.
Sicherheitshinweise
P280 Schutzhandschuhe/Schutzkleidung/Augenschutz/Gesichtsschutz tragen.
P305+P351+P338 BEI KONTAKT MIT DEN AUGEN: Einige Minuten lang behutsam mit Wasser ausspülen.
Eventuell vorhandene Kontaktlinsen nach Möglichkeit entfernen. Weiter ausspülen.