Description
STAMMOPUR RD 5
Instrument cleaner
Concentrate for the removal of contaminants from tools, workpieces and medical devices from production and processing in the dental laboratory.
The preparation is mainly used in dental laboratories to remove machining residues such as waxes and polishing and grinding residues on dental prostheses such as crowns, implants, dentures as well as on machining tools.
Particularly suitable for removing stubborn, dried impurities from blood, secretions, sputum and tissue residues.
The application is carried out both in an ultrasonic bath and in an immersion bath.
Also for manual cleaning and pre-cleaning of dental, surgical, invasive and non-invasive dental instruments, accessories and components of medical devices.
The use in the process of instrument reprocessing takes place in the dental practice, dental clinic, oral surgery and orthodontics.
At a glance
- Concentrate
- Very high material compatibility
- High cleaning effect in the ultrasonic bath
- Removal of stubborn, dried-in contaminants
Material compatibility and impurities
STAMMOPUR RD 5 has a very high material compatibility.
Not suitable for use on aluminium and alkali-sensitive materials such as copper, brass, tin, zinc.
Removes
- Grow,
- Polishing,
- Grinding residues,
- Blood,
- Secretions,
- Sputum
and tissue remnants
From the following materials
- Stainless steel,
- Instrument steel,
- Titanium,
- Glass,
- Porcelain,
- Ceramics,
and plastic.
Application
The exclusively professional users are trained specialists.
The application is intended for cleaning and pre-cleaning as part of technical cleaning in the dental laboratory as well as for manual instrument preparation in the practice and clinic.
Dosage and contact time
Dosage:
in ultrasonic bath: 3%
without ultrasound: 5%
Exposure time:
in ultrasonic bath: 3-10 min
without ultrasound: exposure time depending on the type and stubbornness of the contamination, max. 60 minutes.
Temperature:
For fresh contamination with blood and proteins, prepare cold.
Can be heated up to 60 °C for stubborn contamination.
Composition per 100g:
1-< 10 % anionic surfactants,
1-< 10 % non-ionic surfactants,
1-< 10 % Complexing agent,
1-< 10 % 2-propanol,
1-< 10 % Phenols,
1-< 1.0 % Sodium hydroxide.
Product pH range: alkaline
Application diluted with water of at least drinking water quality.
Additives must not be mixed into the cleaning solution.
Special instructions
Do not treat optics, light guides and bonded parts with ultrasound.
Hazard statements
H315 Causes skin irritation.
H318 Causes serious eye damage.
Precautionary statements
P280 Wear protective gloves/protective clothing/eye protection/face protection.
P305+P351+P338 IF IN EYES: Rinse cautiously with water for several minutes.
Remove contact lenses if possible. Continue to rinse.