Homepage » Products » Laboratory » SONOPULS ultrasonic homogenisers » SONOPULS HD 5000

SONOPULS HD 5000

Homogenisation, deagglomeration, emulsification, suspension, acceleration of chemical reactions or disruption of cells, bacteria, fungi or spores are the main areas of application. Greatly reduced process times and quickly available reproducible results make SONOPULS ultrasonic homogenisers indispensable for modern processes and analytics.

The SONOPULS HD 5020 is ideally suited for gentle sonication of the smallest sample volumes of 0.1-25 ml at 30 kHz with sonotrodes with a diameter of 1.5-2.5 mm. The generator produces a power of up to 20 W.

Bethe ready-to-use set:
for volumes from 0.1 - 25 ml
Ultrasonic nominal power max. 20 W
- Ultrasonic generator GM 5020
- Ultrasonic transducer UW 5020
- Sonotrode MS 1.5, Ø 1.5 mm

The SONOPULS HD 5050 is particularly suitable for gentle sonication of smaller sample volumes of 0.5 -100 ml at 20 kHz and sonotrodes with a diameter of 2-9 mm. The generator operates with a power of up to 50 W.

Ready-to-use set:
for volumes from 0.5 -100 ml
Ultrasonic nominal power max. 50 W
- Ultrasonic generator GM 5050
- Ultrasonic transducer UW 5050
- Sonotrode TS 102, Ø 2 mm

Modern operating concept

Ultrasonic homogenisers fulfil a wide variety of tasks in everyday laboratory work. The range of devices is correspondingly diverse. Understanding the basic design of the homogenisers and the resulting application-related selection of the individual components is the basis for successful application.
0101
Amplitude and power adjustment
Amplitude adjustment in 1-% steps (in the range of 10-100%) for all horns. The alternative power control in watts is also possible via a slider or a numerical input. The display of the actual values allows continuous process control.
0202
Time setting and sequence
Selectable time settings: Timer (countdown) or continuous mode (up to 99 h : 59 min : 59 s). Continuous mode displays the elapsed time, while timer mode displays the remaining time.
0303
Pulsation
For safe sonication of temperature-sensitive samples, the pulse interval can be individually set in 0.5 s steps. The desired sonication duration and the pause can be set independently of each other in the range of 0.5-600 s.
0404
Programme memory
Save recurring processes as a programme in order to start them quickly and conveniently at the touch of a button. at the touch of a button. Up to 8 programmes can be combined and played automatically in any order. automatically one after the other.
0505
Process display
Display and control of all set parameters of the running programme during operation including the remaining time or elapsed time.
0606
Temperature monitoring
The optional TM 5000 temperature sensor ensures constant monitoring of the sample temperature. As desired, a warning signal appears when the limit temperature is reached or the ultrasound is switched off directly.

Ultrasonic transducer

Button
There is a button on the ultrasound transducer. This can be used to switch the ultrasound operation on/off and to control a hand-guided pulsation. The ultrasonic transducer also has a connection socket for using a temperature sensor for monitoring the sample temperature.

Connection for temperature sensor TM 5000
For temperature monitoring, the temperature sensor is connected to the
socket provided for this purpose, which is otherwise covered with a dust cap. A temperature display appears on the generator, enabling the user to record the temperature at any time. If the limit temperature is exceeded, a warning signal sounds and/or the process is automatically stopped.

Ultrasound generator

Ultrasound generator in easy-care, robust plastic housing with connection for all SONOPULS ultrasound transducers of the HD 5000 series as well as a recessed grip for easy transport and adjustment on the laboratory bench. The modern 7″ touch display offers intuitive, user-friendly operation. The setting of the target values for amplitude, pulsation and time and the display of the actual values enable reproducibility of the results. reproducibility of the results.

All information in the

Product flyer

Ihr Browser unterstützt das Einbetten von PDF-Dateien nicht. Sie können die Datei here download.

Public address vessels for direct sound reinforcement

Rosette cells RZ
The rosette cells allow an even and intensive sonication of liquid media. Due to the sound pressure, the sample is pressed against the bottom of the vessel and thus through the three side arms and can circulate well. This results in continuous mixing of the medium. When the rosette cells are placed in an ice bath, the contents are effectively cooled due to the enlarged glass surface and the good circulation.

Cooling vessels KG
Sonication results in the conversion of mechanical energy into heat (due to internal friction in the liquid) and thus in a more or less high heating of the samples. Cooling of the medium may therefore be necessary for temperature-sensitive samples. The sample vessels can be placed in an ice bath, for example. However, the immersion depth of the sonotrode is not visible this way. A better alternative is the KG cooling vessels with cooling jacket for connection to an external cooler. This enables controlled temperature control during sonication.

 

Flow vessels DG
With cooling jacket. Continuous sonication of samples up to 30 l/h in flow is possible. The cooling jacket allows temperature control by liquid coolant during sonication.

Insulation

Noise reduction by approx. 30 dB (AU)

cleaning

Easy-to-clean & removable drip tray and insert In the interior made of stainless steel

New: Sound proof box LS 40

Cavitation produces noise which is perceived as unpleasant by the user and others in the vicinity. To reduce the noise level, it is recommended to use a noise box.

Process control

LED interior lighting and acrylic glass window for process monitoring

Accessories

Sufficient space for direct and indirect applications by using the holding frame HG 40 and further optional accessories

Example application in the video

We have put together an overview of sample cleanings to help you carry out optimal ultrasonic cleaning. Please visit our Youtube channel for more videos.

further

Accessories

Generators

The ultrasonic generator converts the low-frequency mains voltage into a high-frequency voltage of 20 kHz.

Generators

learn more

Step & Booster Horns

A step or booster horn amplifies the amplitude generated by the ultrasonic transducer with its concentrating design.

Step & Booster Horns

To the accessories

Sonotrodes

They transmit the mechanical vibrations into the sample.

Sonotrodes

learn more

Holding frames

The holding racks are used for the correct and variable positioning of the ultrasound transducers.

Holding frames

To the accessories

Ultrasonic transducer

An ultrasonic transducer (UW) is used to convert the electrical energy supplied by the ultrasonic generator into mechanical vibrations.

Ultrasonic transducer

To the accessories

Public address system

Various sound vessels for direct & indirect sound reinforcement.

Public address system

To the accessories

Bandelin exhibition stand

Visit us!

We are represented at numerous trade fairs. Visit us at our stand and let us advise you on your ultrasonic application. You can find the next trade fair dates where we are represented in the overview. Please feel free to contact us and arrange an appointment at the trade fair for your request. We are looking forward to your visit!

This will close in 0 seconds

EN