Description
TICKOMED 1
Dental Instrument Cleaner
Concentrate for manual cleaning and pre-cleaning of dental, surgical, invasive and non-invasive medical instruments, rotating and non-rotating small and precision instruments as well as small endodontic instruments, accessories and components of medical devices, both in an ultrasonic bath and in an immersion bath process.
The preparation is used in the process of instrument preparation in dental practices, dental clinics, oral surgery and orthodontics.
The application is intended for cleaning and pre-cleaning as part of manual reprocessing.
At a glance
- Concentrate
- Removes blood, secretions and saliva, bone and
Tooth substance, dentin
- Very high material compatibility, also for
Light metals
- Can be used as a contact liquid
- Biodegradable
- Very high cleaning effect in the ultrasonic bath
Material compatibility and impurities
TICKOMED 1 has a very high material compatibility.
Due to the formulation in the mild-alkaline range, the preparation is even suitable for aluminium and anodised parts.
Removes fresh and stubborn contaminants, such as
- Blood,
- Secretion,
- Spoke
- Dentine,
- Bone
and tooth structure
From the following materials
- Metal,
- Aluminium,
- anodised parts,
- Titanium,
- Glass,
- Porcelain,
- Ceramics
and plastic.
Application
The exclusively professional users are trained specialists.
TICKOMED 1 has a very high cleaning effect. In the ultrasonic bath, it leads to a reduction in the time of the cleaning process due to the physical support in the multitude of applications.
Recommended as contact liquid for the indirect process in the ultrasonic bath for STAMMOPUR GR and STAMMOPUR Z.
Dosage and contact time
Dosage:
in ultrasonic bath: 3%
without ultrasound: 5%
Exposure time:
in ultrasonic bath: 3-10 min
without ultrasound: exposure time depending on the type and stubbornness of the contamination, max. 12 hours. For aluminium and its alloys, do not exceed an exposure time of 60 minutes.
Temperature:
For fresh impurities, prepare cold,
Can be heated up to 60 °C for stubborn contamination.
Composition per 100g:
5-15 % non-ionic surfactants,
5-15 % Cationic surfactants,
< 5 % Phosphate,
Complexing agent
Product pH range: mildly alkaline
Application diluted with water of at least drinking water quality.
Additives must not be mixed into the cleaning solution.
Special instructions
Do not treat optics, light guides and bonded parts with ultrasound.
Hazard statements
H318 Causes serious eye damage.
H315 Causes skin irritation.
Precautionary statements
P280 Wear protective gloves/protective clothing/eye protection/face protection.
P305+P351+P338 IF IN EYES: Rinse cautiously with water for several minutes.
Remove contact lenses if possible. Continue to rinse.