Description
STAMMOPUR DR 8
Instrument disinfection and intensive cleaning
Concentrate for manual, chemical disinfection and disinfectant, non-fixing intensive cleaning of general, surgical, invasive and non-invasive medical instruments and accessories in ultrasonic baths as well as in immersion baths.
The preparation is used in the process of instrument reprocessing in the AEMP (reprocessing unit for medical devices) in the clinic, the doctor's and dentist's practice, as well as in the dental laboratory and in medical foot care.
For critical and semi-critical medical devices, further reprocessing steps are necessary, do not use for final disinfection.
Spectrum of action
- Bactericidal and levurocidal (EN 13727, EN 145661, EN 13624, EN 14562, DGHM, high load)
- Limited virucidal (EN 14476, high load)
At a glance
- Concentrate
- VAH listed
- Economical due to low application concentration
- Very high material compatibility
- Free from aldehydes, chlorine and phenols
- Bactericidal, levurocidal, limited virucidal
- Effective against coronaviruses (SARS-CoV-2)
- Biodegradable
- Very high cleaning effect in the ultrasonic bath
Material compatibility and impurities
STAMMOPUR DR 8 has a very high material compatibility.
Due to the formulation in the mild-alkaline range, the preparation is also suitable for light metals within the scope of the exposure times.
Cleans manually, disinfecting, non-fixing and intensively from the following materials
- Metal,
- Light metals,
- Titanium,
- Glass,
- Porcelain,
- Ceramics
and plastic.
Application
The exclusively professional users are trained specialists.
STAMMOPUR DR 8 has a very high cleaning effect. In the ultrasonic bath, it leads to a time reduction of the cleaning process in the multitude of applications. The time reduction of the disinfection due to the ultrasonic effect, which has been confirmed by an expert, additionally leads to an overall shortening of the reprocessing process.
For endoscopes and accessories, the suitability recommendations of the companies Karl Storz GmbH and Richard Wolf GmbH are available. The instructions of the instrument manufacturers must be strictly observed when reprocessing with STAMMOPUR DR 8, especially in the ultrasonic bath.
STAMMOPUR DR 8 is not compatible with aldehydic preparations. These must be thoroughly removed from cleaning materials and accessories before using STAMMOPUR DR 8.
The standing time of the disinfectant solution has been extended to three days by expert opinion.
Dosage and exposure times
Application with ultrasound:
Bactericidal, levurocidal,
limited virucidal incl. H5N1: 2% - 5 min
SV40 2 % - 10 minutes
Adeno 3 % - 15 minutes
Application without ultrasound:
Bactericidal, levurocidal: 1 % - 60 min
Bactericidal, levurocidal, limited virucidal
incl. H5N1 and additionally against SV 40: 2 % - 30 min or 3 % - 15 min
Temperature: prepare cold and do not heat additionally (max. 45 °C).
Adeno (EN 14476, high load) Effectiveness only when used in an ultrasonic bath.
Composition per 100g:
9.9 g bis(3-aminopropyl)dodecaylamine,
8.4 g didecyl methyl polyoxyethyl ammonium propionate,
< 30 % butane-1,4-diol,
5-15 % butyl diglycol,
5-15 % non-ionic surfactants,
< 5 % Piperazine,
Complexing agents,
pH regulators.
Product pH range: mildly alkaline
The application is diluted with water of at least drinking water quality.
Additives must not be mixed into the solution.
Special instructions
Do not treat optics, light guides and bonded parts with ultrasound.
Hazard statements
H302 Harmful if swallowed.
H314 Causes severe skin burns and eye damage.
H317 May cause an allergic skin reaction.
H334 May cause allergy, asthma-like symptoms or breathing difficulties if inhaled.
breathing difficulties.
H336 May cause drowsiness and dizziness.
H361fd Suspected of damaging fertility. Suspected of damaging the unborn child.
in the womb.
H411 Toxic to aquatic life with long lasting effects.
Safety advice
P280 Wear protective gloves/protective clothing/eye protection/face protection.
P305+P351+P338 IF IN EYES: Rinse cautiously with water for several minutes.
Remove contact lenses if possible. Continue to rinse.
P308+P313 IF EXPOSED or if affected: Get medical advice/attention.
P310 Immediately call a POISON CENTRE/doctor.